Electrode calibration procedure – Hanna Instruments HI 84529 User Manual

Page 25

background image

www.

hannainst

.com

25

Operational Guide

Procedure

A single one, two or three-point calibration can be performed, using the four

predefined buffers 4.01, 6.00, 8.30 and 10.01 pH. For a single point calibration

any of the four buffers may be used, 8.30 pH is recommended.

Note: The HI 84529 will not accept other pH buffers for calibration.

1. Pour small quantities of selected buffer solutions into clean beakers. For

accurate calibration used two beakers for each buffer solution, the first

one for rinsing the electrode and the second one for calibration.

2. Put a magnetic stir bar in the beaker that will be used for calibration.

3. Remove the protective cap and rinse the electrode with some of the

buffer solution to be use for the first calibration point.

4. Put the first beaker with calibration buffer in the beaker holder.

5. Place the electrode holder on the top of the beaker and secure it by

turning clockwise and press STIR.

6. Immerse the pH electrode and the temperature probe approximately

2 cm (0.8”) into the buffer, paying attention not to touch the stir bar.

To enter Electrode Calibration follow the next steps:

1. Press CAL function key then ELECTRODE.

2. The electrode calibration screen will be displayed.

3. Press CLEAR to delete the previous calibration.

1 Point Calibration

1. The 8.30 buffer will be selected

by default. If necessary, press the

ARROW keys in order to select a

different buffer value.

2. The

(unstable measurement)

symbol will be shown on the display

until the reading becomes stable.

3. When the reading is stable and

close to the selected buffer, the

(unstable measurement) symbol will

disappear and the CONFIRM key will

become active.

4. Press CONFIRM to confirm the

Electrode

Calibration

Procedure