Operational Guide

For best accuracy, before taking any measurement, ensure that the pump is

calibrated on the selected range following the “Pump Calibration Procedure”.

Note: Verify that the instrument has been calibrated (pH and pump) before

performing any titrations.

Refer to “Setup Menu” to set up the instrument for your measurement. Select

the corresponding measurement range.

By Mass:

Place a clean 100 mL plastic beaker on an

analytical balance. Zero the balance.

Add 20 ± 1 g (Low 20, High 20) or 50 ± 1 g

(Low 50) to the beaker.

Record the reading once it has stabilized.

Remove the beaker from the balance.

By Volume:

Volumetrically add 20 mL( Low 20, High

20) or 50 mL (Low 50) to a clean 100mL

plastic beaker.

Note: Failure to use a clean pipette will

result in erroneous readings.

Note: DO NOT PLACE THE TIP INTO THE

SAMPLE BEAKER. PLACE THE TIP OVER

A WASTE BEAKER. A SMALL AMOUNT OF

TITRANT IS DISPENSED WHEN THE PUMP

RESETS.

Press START to begin a titration.

For samples by mass the “Edit Sample

Mass” screen will be displayed. Use the

ARROW keys to enter the exact sample

mass then press ACCEPT.

Titration

Procedure

Unit

Low Range 20 and 50

High Range 50

%l.a

0.01 to 0.20

0.1 to 2.0

°SH

0.4 to 8.9

4.4 to 88.9

°D

1.0 to 20.0

10 to 200

°TH

1.1 to 22.2

11.1 to 222.2