Operational Guide

If CAL DUE is displayed perform an

electrode calibration.

Options:

Press MENU to access the instrument’s

menu.

Press HELP to view the contextual help

whenever time you need additional

information.

Press STIR to start/stop the stirrer.

Press TITRATOR to enter titration mode.

Press CAL to access the calibration menu.

Press LOG to save the current reading. A

message indicating the free log space will

be displayed for a few seconds.

In order to take pH measurements,

follow the next steps: Submerge the pH

electrode 2 cm (0.8’’) and the temperature

probe into the sample to be tested and

stir gently. Allow time for the electrode

to stabilize. When the reading becomes

stable, the (unstable measurement)

symbol will disappear.

If the pH reading is less than -2.00 pH or

greater than 16.00 pH, the closest full-

scale value will be displayed blinking.

If measurements are taken successively

in different samples, it is recommended

to rinse the electrodes thoroughly with

deionized or distilled water and then

with some of the next sample to prevent

cross-contamination.

The pH measurements are affected by temperature. In order to have accurate

pH measurements, the temperature effect must be compensated for. To use the

Automatic Temperature Compensation (ATC) feature, connect and submerge

the HI 7662-T temperature probe into the sample as close as possible to the

electrode and wait for a few seconds. The “ATC” message will be shown on the

pH