For 20 g and 20 mL samples add deionized

or distilled water to the 60 mL mark.

Place the stir bar in the beaker and put

the beaker in the mini titrator top.

Place the probe holder on the top of the

beaker and secure it by turning clockwise.

Rinse the pH electrode, reference electrode

and temperature probe with deionized

water and immerse it roughly 1.5 cm (0.6”)

into the sample. Be sure that the tip of the

electrode is not hitting the stir bar. Insert

the dosing tip into the titrant tube sleeve.

IT IS CRITICAL THAT THE TIP BE IMMERSED

APPROXIMATELY 0.25 CM (0.1") INTO THE

SOLUTION BEING TITRATED.

Press CONTINUE to begin the titration

and STOP to abort it.

The instrument will continuously update

the concentration on the display. The

value will be displayed blinking. When

the reading is under range “----” symbol

appears blinking.

The titration curve can be visualized

during a titration by pressing Plot ON.

Press Plot OFF or ESC to exit this mode.