Operational Guide

NOTE: DO NOT PLACE THE DISPENSING

TIP INTO THE CALIBRATION BEAKER,

PLACE THE TIP OVER A WASTE BEAKER.

A SMALL AMOUNT OF TITRANT IS

DISPENSED WHEN THE PUMP RESETS.

Press START, wait for the syringe to refill.

Place the stir bar in the beaker and put

the beaker in the minititrator top.

Place the probe holder on the top of the

beaker and secure it by turning clockwise.

Rinse the pH electrode, reference electrode

and temperature probe with deionized

water and immerse them roughly 1.5 cm

(0.6”) into the sample. Be sure that the tip

of the electrode is not hitting the stir bar. If

necessary, additional distilled or deionized

water can be added.

Insert the dosing tip into the titrant tube

sleeve. IT IS CRITICAL THAT THE TIP BE

IMMERSED APPROXIMATELY 0.25 CM (0.1")

INTO THE SOLUTION BEING TITRATED.

Press CONTINUE to begin the calibration

and STOP to abort it. At the end of the

calibration, ”Calibration Completed“

appears on display. To repeat the calibration

press Restart or ESC to return to the main

screen.

Note: If temperature probe is not

connected, Manual Temperature

Compensation is used and MTC appears

on the right side of the screen. If

Automatic Temperature Compensation is

in use the ATC appears on the right side

of the screen.

Pump

Calibration

Procedure