beautypg.com

Hanna Instruments HI 4211 User Manual

Page 58

background image

5 8

• If accessing Log History option while logging, last logged data can be visualized on the LCD (see Display

Mode section for details, page 11).

• To stop the logging session, press

or

/

. The Log Save screen will be displayed in which the

logged lot parameters can be set:

• Press

to adjust the log interval and/or the log

sampling or press

to save the current log in

the displayed format.

• Press

to enter log interval edit menu and use

or

to adjust the logging start/stop

time or the log sampling. Press

to save the

current value and use

or

to adjust

next/previous parameter.

Note: The minimum log sampling which can be set is calcu-

lated related to the logging time thus the 5000 records/lot
not to be exceeded (e.g. for a 50.000 seconds logging
time, the minimum log sampling will be 10 seconds).

• Press

to exit log interval edit menu and then

press

to save the current log with the new

configuration. While the instrument is saving the data,
a “Please wait...” pop-up message will be displayed
on the LCD.

Note: For larger logged lots, the saving time can be up to

several minutes.

LOGGING MODE 2

This logging mode can be used for multiple samples measurement. By choosing this logging mode,

and

will be available in

Measure mode.

Note: For HI 4212, the

or

and

or

will be available in multichannel

Measure

mode, depending on the focused channel.

To log data using this mode:
• Press

while in

Measure mode to start the logging session. When the measured value is frozen on

the LCD by pressing

and the stability criterion is reached, the logged value is the one that has been

frozen on the LCD until returning to normal logging mode by pressing

.