Hanna Instruments HI 4211 User Manual

Page 34

background image

3 4

Analyte Addition
This method is similar to the Known Addition method, with the difference that an aliquot of sample is added to a
standard of known concentration. The sample and standard contain the same ion to be measured. The ion
concentration is then calculated using the difference in mV potential.
Analyte Subtraction
In the Analyte Subtraction method an aliquot of sample is added to a standard of known concentration, reacting
with the ion to be measured. The stoichiometric ratio between standard and sample must be known. The ion
concentration is then calculated using the difference in mV potential. The method is especially usefull when there
is no available ISE for the measured ion.

To set the Reading Mode:
• Press

while in

ISE Measure mode.

• Press

.

• Use

or

to select the Reading Mode option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.

Temperature – See pH Setup section, page 20.
Notes: • To access ISE Setup, press

and then

.

• The Temperature Compensation option is also included in the Temperature parameter.

Temperature Compensation
This option allows the user to enable / disable the temperature
compensation for the measurement of Ion concentration.

To set the Temperature Compensation:
• Press

while in

ISE Measure mode.

• Press

.

• Use

or

to select the Temperature option.

• Press

and use

or

to highlight the

Temperature Compensation option.

• Press

and use

or

to enable / disable

the Temperature Compensation.