Hanna Instruments HI 4211 User Manual

Page 38

background image

3 8

Isopotential Point
This option allows the user to edit the isopotential point of the electrode used for ion measurements. The ion
selective electrodes have different isopotential points. If temperature compensation is desired for the ion
measurements, the isopotential point value is absolutely necessary.
The isopotential point is always edited in ppm (mg/L) units.

To set the Isopotential Point:
• Press

while in

ISE Measure mode.

• Press

.

• Use

or

to select the Isopotential Point option.

• Press

and use

or

to increase / decrease

the isopotential point value.

• Press

key to save the modified value.

• Press

to return to the Ion Setup options. If the Saving

Confirmation is enabled, press

to accept the modified option,

to escape without saving or

to return to the editing

mode. Otherwise, the modified option is saved automatically.

Note: If an ISE calibration was performed and the Isopotential Point option is changed, a warning message

appears on the LCD informing the user to perform a new calibration or to set the previous option in order
to perform accurate measurements.

ISE Significant Digits
Accessing this option, the number of ISE significant digits can be set, with one (x), two (xx) or three (xxx)
significant digits.

To set the ISE Significant Digits:
• Press

while in

ISE Measure mode.

• Press

.

• Use

or

to select the ISE Significant Digits

option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.