background image

4 7

Two standard entry types are available: ManualSelection and Custom Standard. The default option is Manual Selection.

A. To calibrate the instrument using Manual Selection standard entry type:
• Press

. If the instrument was calibrated before and calibration was not cleared, the old calibration can be

cleared by pressing

. After 10 seconds,

will no longer be available.

Note: It is very important to clear calibration history when a new electrode is used

because most errors and warning messages that appear during calibration
depend on calibration history.

• Add ISAB to all standard solutions.
• Immerse the Ion Selective Electrode and the temperature probe approximately 4

cm (1½”) into the less concentrated standard solution and stir gently.

• Select the used standard solution concentration with

or

. The

“Please wait...” message will appear on the LCD until the reading is stable or
the concentration is validated.

Note: When in ATC mode, if pressing

a pop-up menu will be displayed on the LCD in which the

concentration value can be ajusted using

or

. Press

to save the new

concentration value.

• If the standard solution concentration is validated,

will appear on the LCD. Press

to update

calibration. The calibration point value will be added to the Calibrated Standards section.

• The “Please wait...” message will appear on the LCD for 10 seconds and until the reading is stable, time in

which immerse the ion selective electrode and the temperature probe into the next standard solution and
follow the above procedure or press

to exit calibration.

B. To calibrate the instrument using Custom Standard entry type:
• Press

. If the instrument was calibrated before and calibration was not cleared, the old calibration

can be cleared by pressing

. After 10 seconds,

will no longer be available.

Note: It is very important to clear calibration history when a new electrode is

used because most errors and warning messages that appear during
calibration depend on calibration history.

• Add ISAB to all standard solutions.
• Immerse the Ion Selective Electrode and the temperature probe approximately 4

cm (1½”) into the less concentrated standard solution and stir gently.

• When the stability condition is reached and 10 seconds elapsed, a pop-up menu

will appear on the LCD in which the standard concentration value can be set.

• Press

/

to select next/previous entry to be edit. Press

and use

or

to set the desired value. Press

to save the modified value and then press

to

store the set concentration value.