the user to select the desired character. It is also possible to delete the last character by positioning the cursor on
the Backspace character and pressing

.

• Press

to return to Sample ID options. If the Saving

Confirmation is enabled, press

to accept the modified

option,

to escape without saving or

to

return to the editing mode. Otherwise, the modified options
are saved automatically.

• If the selected increment mode is Automatic, the desired sample ID

value can be set using

or

.

• Press

to save the current value or press

to

cancel operation.

Stability Criteria
This option allows the user to select the signal stability criterion for the measured parameter (pH, mV, ISE):
Fast – this criterion will give faster results with less accuracy.
Medium – this criterion will give medium speed results with medium accuracy.
Accurate – this criterion will give slower results with high accuracy.

To set the Stability Criteria:
• Press

while in

pH Measure mode.

• Press

.

• Use

or

to select the Stability Criteria option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.

Reading Mode
This option allows the user to select between Direct and Direct/AutoHold pH reading mode. If choosing the second
option, the current reading can be frozen on the LCD when

is pressed and if the stability criterion is reached.