• Use

or

to select the Log option.

• Press

and use

or

to highlight the

New Lot option.

• Press

to generate a new manual lot. A pop-up menu

will be displayed asking for confirmation.

• Press

to confirm or press

to escape without

saving and return to the Log options.

Alarm
This option allows the user to edit the alarm settings: Alarm State and Alarm Limits. If the Alarm option is enabled,
a continuous double beep will be heard, along with the “Alarm” indicator blinking on the LCD, each time the set
limits in

Measure mode are exceeded.

Alarm State
Three modes are available for the Alarm State option:
Disabled – the alarm will be disabled.
Inside Limts – the alarm will notify the user when the measured value is inside the set limits.
Outside Limits – the alarm will notify the user when the measured value is outside the set limits.

To set the Alarm State:
• Press

while in

pH Measure mode.

• Press

.

• Use

or

to select the Alarm option.

• Press

and use

or

to highlight the

Alarm State option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.

Alarm Limits
This option allows the user to set the alarm limits for the measured value.
Note: The Alarm High value can not be lower than the Alarm Low value.