mV & Relative mV MEASUREMENTS

mV/ORP MEASUREMENTS

Oxidation-reduction potential (ORP) measurements provide the quantification of the oxidizing or reducing power of the
tested sample.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.

DIRECT MEASUREMENT

To measure the mV of a sample using the Direct reading mode:

• Press

and then

to enter

mV Measure mode

(for the selected channel – HI 4212 only).

• Select the Direct reading mode (see mV Setup for details, page 32).
• Submerge the tip of the ORP electrode (4 cm/1½”) into the

sample to be tested and allow a few seconds for the reading
to stabilize.

• The instrument will display the measured mV value on the LCD.
Note: If the reading is out of range, “-----” will be displayed

on the LCD.

DIRECT/AUTOHOLD MEASUREMENT

To measure mV of a sample using the Direct/AutoHold reading mode:

• Press

and then

to select

mV Measure mode

(for the selected channel – HI 4212 only).

• Select the Direct/AutoHold reading mode (see mV Setup for

details, page 32).

• Submerge the tip of the ORP electrode approximately 4

cm (1½”) into the sample to be tested and allow a few
seconds for the reading to stabilize.

• The measured mV value will be displayed on the LCD. If

pressing

, the “AutoHold” indicator will start blinking

on the LCD until the stability criterion is reached. The mV value
will be frozen on the LCD, along with “AutoHold” indicator.

• To return to normal

Measure mode press

.

Note: If the reading is out of range, “-----” will be displayed

on the LCD.