beautypg.com

Chapter 7 - flow meter calibration, Method 1 – known volume, Step 1 – known value – TeeJet 844-E Sprayer Control User Manual

Page 33: Step 2 – programming calibration number, Step 3 – resetting volume counter, Step 4 – spraying known volume, Step 6 – double-checking the new value, Method 2 – known tip size method, Step 1 – check tip size, Step 2 – count tips

background image

29

98-70006-ENUS R4

844-E Sprayer Control

ChaptEr 7 - Flow MEtEr Calibration

The flow meter supplied with your system has been calibrated at the factory and under normal circumstances there may be no need to re-calibrate

it. However, the factory calibration setup may not reflect specific sprayer plumbing. Before spraying actual chemicals, the flow meter should be

checked for proper calibration. Also, the flow meter is made up of moving parts and can wear over time. If you feel that your flow meter is worn,

you can re-calibrate your flow meter using the 844 console. To do so, use either of the following methods:

Method 1 – Known Volume

Step 1 – Known Value

Select a known volume of water (n) that you will pump through the flow meter. In these steps, 10 US gallons (38 liters) {8 Imperial gallons} will be

used as an example. The specific volume you use is not important, although larger volumes generally produce more accurate results.

Step 2 – Programming Calibration Number

Go into the System Setup mode and advance to the flow meter calibration number step (flashing turbine at the top of the display). In this

programming step, enter (n) 10 US gallons (38 liters) {8 Imperial gallons} as the calibration number. Exit the System Setup mode by pressing and

holding the key for five seconds.

Step 3 – Resetting Volume Counter

Reset the total volume number that is displayed in the lower right portion of the display. (Display alternates between Ac (Ha) covered and the

volume sprayed through the flow meter). To reset this number, press and hold the Plus key and Minus key simultaneously for three (3)

seconds or until you see the values reset to zero.

Step 4 – Spraying Known Volume

Engage the pump on the sprayer, flip the Master boom section of the 844 to “ON” (make sure at least one boom section switch is on), and pump

exactly (n) 10 US gallons (38 liters) {8 Imperial gallons} through the flow meter. When the known volume is reached, shut the Master boom switch

“OFF”.

Step 5 – Entering in New Flow meter Calibration Number

The number that is displayed in the lower right corner of the display is your new flow meter calibration number. Record this number, enter the

System Setup mode and advance to the flow meter calibration number step. Enter the new value from your test into this step and then exit the

System Setup mode.

Step 6 – Double-checking the New Value

Double-check your new value by spraying a larger volume of water, such as 100 US gallons (378 liters){83 Imperial gallons} through the flow

meter. Before spraying the larger volume, make sure to reset the memory counters by depressing and holding the Plus key and Minus key

simultaneously for three (3) seconds.
NOTE: After double checking the new calibration number, if you find that the volume shown on the display of the 844 is more than +/- 3% in error,

change the flow meter calibration number in the System Setup mode by the % difference it is in error.

Method 2 – Known tip Size Method

Step 1 – Check Tip Size

Determine what size tip is on the sprayer. An actual flow collection should be done on a representative sample to ensure the tips are not worn. To

do this, set the boom pressure at 40 psi (2 bar). This must be pressure at the tips (i.e. use manual pressure gauge mounted on the boom). Collect

flow for 1 minute from 1 tip.
Example: xR8004 tip should spray 0.4 gallons in 1 minute (1.29 LPM) {0.32 Imp gal}

Step 2 – Count Tips

Count the total number of tips to be used for measure. You can use 1 boom section, 2 boom sections, etc.
Example: Boom Section 1 has 10 tips

Step 3 – Calculated Flow

Multiply the number of tips to be measured by the size (or individual flow rate) of each tip.
Example: 10 tips x 0.4 gPM (1.29 LPM) {0.32 Imp gPM} = 4gPM (12.9 LPM) {3.2 Imp gPM}