background image

Perform external adjustment / function "CAL 01":
 Call up function "CAL 01”

 Press the TARE button, the currently set adjustment weight is displayed. The last

digit flashes.

To change the adjustment weight select the digit to be changed with BG/ NET
key, the active digit flashes. Enter the desired value with the navigation keys.
Press the TARE key repeatedly, until "-----“ is indicated.

 Ensure that there are no objects on the weighing pan. Press the TARE key.

The weight value of the required adjustment weight will be displayed.

98