Perform external adjustment / function "CAL 01":
Call up function "CAL 01”
Press the TARE button, the currently set adjustment weight is displayed. The last
digit flashes.
To change the adjustment weight select the digit to be changed with BG/ NET
key, the active digit flashes. Enter the desired value with the navigation keys.
Press the TARE key repeatedly, until "-----“ is indicated.
Ensure that there are no objects on the weighing pan. Press the TARE key.
The weight value of the required adjustment weight will be displayed.