8 applications, 1 parts counting – KERN & SOHN KXS-TM User Manual
Page 21
8 Applications:
8.1 Parts counting
Before the balance can count parts, it must know the average part weight (i.e.
reference). Proceed by putting on a certain number of the parts to be counted. The
balance determines the total weight and divides it by the number of parts, the so-
called reference quantity. Counting is then carried out on the basis of the calculated
average piece weight.
As a rule:
The higher the reference quantity the higher the counting exactness.
Determine the unit weight and count
1. In the weighing mode press the unit key, in order to activate the counting
function. The indicator Pcs is indicated.
2. Press the BG/NET-key repeatedly until the desired reference number of items
(e.g. 100) appears, select between C 10, C 20, C 50, C100, C 200.
3. Place as many items to be counted (such as 10 items) as demanded by the
set reference quantity and confirm by UNIT key. The weighing scales calculate
the reference weight. The current number of items (e.g. 10 pieces) is
indicated.
4. Remove reference weight. The balance is from now in parts counting mode
counting all units on the weighing plate.
KXS/KXG-TM-BA_IA-e-1420
21