2 linearization/ menu block – KERN & SOHN KXS-TM User Manual
Page 93
19.2 Linearization/ menu block
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range. If linearity deviation is discovered during a monitoring of test resources, you
can improve this by means of linearization.
• Carrying out linearization is restricted to specialist staff possessing
well acquainted with the workings of weighing scales.
• The test weights to be used must be adapted to the weighing scale’s
specifications; see chapter “monitoring of test resources”.
• Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
• After successful linearisation you will have to carry out calibration;
see chapter 3.4 “Monitoring of test resources”.
Procedure:
1. Call up function
, see chapter 18.
2. Press the TARE key.
Ensure that there are no objects on the weighing pan.
3. Press the TARE key.
4. With display "L1 1" set the „first adjusting weight carefully into the centre of the
weighing plate". With the navigation keys (see chapter 2.1.1) Enter the desired
factor, by which the following adjustment weight is to increase.
Press the TARE key.
5. With display "L2 1" set the second adjustment weight carefully into the centre of
the weighing plate. With the navigation keys (see chapter 2.1.1) Enter the desired
factor, by which the following adjustment weight is to increase.
Press the TARE key.
KXS/KXG-TM-BA_IA-e-1420