Example 530010:
5
⇒
Readability of 5 steps
3
⇒
3 decimal places d = 0.005
0
⇒
Standard weighing unit „kg
“
0
⇒
Single-range balance
1
⇒
Sequence of the weighing units
0
⇒
Multi-interval balance
Press the TARE key, the newly set capacity / resolution is displayed for 2 seconds
(display sample maximum 15 kg / d 0.005kg). The display automatically returns
back to the menu.
Select „CSP 00“ with the navigation keys.
Confirm with TARE key.
Exit menu / jumping back to weighing mode:
Move the adjusting switch to the position „LOCK“, see chapter 18.
After the configuration linearize and/or adjust the weighing system.