KERN & SOHN KXS-TM User Manual
Page 95
Setting "factor”
1 =
All weight values are equal
2 =
The following weight value is twice as high as the first.
3 =
The following weight value is three times higher than the first
4 =
….5 times…
5 =
….6 times…
6 =
….7 times…
7 =
….8 times…
8 =
….9 times…
A =
….10 times…
B =
….11 times…
C =
….12 times…
D =
….13 times…
E =
….14 times…
F =
….15 times…
Examples:
Example 1: 4-point linearization/ 30 kg (10 kg, 10 kg, 10 kg)
Display Button
Explication
03 CLn
TARE
Linearization function
L0
TARE
Zero point
L11
TARE
First adjustment weight (10 kg)
L21
TARE
Second adjustment weight (10 kg)
L31
TARE
Third adjustment weight (10 kg)
L41
BG/NEt
Terminate linearization (4-point linearization)
03 CLn
Example 2: 5-point linearization/ 30 kg (5 kg, 10 kg, 10 kg, 5kg)
Display Button
Explication
03 CLn
TARE
Linearization function
L0
TARE
Zero point
L11
TARE
First adjustment weight (5 kg)
L22
TARE
Second adjustment weight (10 kg); factor 2, i.e. twice as high as the first
L32
TARE
Third adjustment weight (10 kg); factor 2, i.e. twice as high as the first
L41
Fourth adjustment weight (5 kg), factor 1, i.e., the same size as the first
L51
BG/NEt
Terminate linearization (5-point linearization)
03 CLn