KERN & SOHN KXS-TM User Manual
Page 106
![background image](/manuals/697288/106/background.png)
19.5.3 Resetting to zero when switched on (function "CFn 04
“
)
Call up function „CFn 04“, see chapter 18.
Press the TARE-key, current setting appears.
With the navigation keys (see chapter 2.1.1) select the desired setting.
0
⇒
Zero setting when switching on within
± 10 % Max.
Factory setting verified models (CFn 02 = 1)
1
⇒
Zero setting when switching on within ± 20 % Max.
2
⇒
Zero setting when switching on within ± 30 % Max.
3
⇒
Zero setting when switching on within ± 40 % Max.
4
⇒
Zero setting when switching on within ± 50 % Max.
5
⇒
Zero setting when switching on within ± 60 % Max.
6
⇒
Zero setting when switching on within ± 70 % Max.
7
⇒
Zero setting when switching on within ± 80 % Max.
8
⇒
Zero setting when switching on within ± 90 % Max.
9
⇒
Zero setting when switching on within ± 100 % Max.
Factory setting unverified models (CFn 02 = 0)
Confirm with the TARE key, the display returns back to the menu.
Select "CFn 00" with the navigation keys.
Confirm with TARE key.
To exit the menu / return back to the weighing mode, see chapter 18.1.1 and/or
18.1.2.
106
KXS/KXG-TM-BA_IA-e-1420