beautypg.com

6 data hold functions (function "fnc 06“) – KERN & SOHN KXS-TM User Manual

Page 40

background image

10.6 Data Hold functions (function "FnC 06

)

With verified balancing systems (Cfn 02 = 1) the setting „hold = 0

cannot be

altered with the function „FnC 06

(changes are possible with function "CFn 05" see chapter 18.5.4.).


 Call up function "FnC 06", see chapter 10 "Navigation in the menu“.

 Press the TARE-key, current setting appears.
 Select desired settings with TARE or UNIT key.

hold 0

Function switched off (factory setting)

hold 1

Peak value function
This function indicates the highest load factor (peak value) of a
continuously rising load. The peak value remains in the display until it
is deleted with any key.

hold 2

„Stable hold 1“ mode
The weight value is held automatically after reaching a stable value
up to the manipulation of any key in the display.

hold 3

„Stable hold 2“ mode
The weight value is held in the display after reaching a stable value
until the load falls under 10d.

hold 4

Animal weighing
This function is suitable for jerky weighing procedures, see following
chapter 10.6.1

40

KXS/KXG-TM-BA_IA-e-1420