6 data hold functions (function "fnc 06“) – KERN & SOHN KXS-TM User Manual
Page 40
10.6 Data Hold functions (function "FnC 06
“
)
With verified balancing systems (Cfn 02 = 1) the setting „hold = 0
“
cannot be
altered with the function „FnC 06
“
(changes are possible with function "CFn 05" see chapter 18.5.4.).
Call up function "FnC 06", see chapter 10 "Navigation in the menu“.
Press the TARE-key, current setting appears.
Select desired settings with TARE or UNIT key.
hold 0
Function switched off (factory setting)
hold 1
Peak value function
This function indicates the highest load factor (peak value) of a
continuously rising load. The peak value remains in the display until it
is deleted with any key.
hold 2
„Stable hold 1“ mode
The weight value is held automatically after reaching a stable value
up to the manipulation of any key in the display.
hold 3
„Stable hold 2“ mode
The weight value is held in the display after reaching a stable value
until the load falls under 10d.
hold 4
Animal weighing
This function is suitable for jerky weighing procedures, see following
chapter 10.6.1
40
KXS/KXG-TM-BA_IA-e-1420