The FT-1x weight indicators have been designed as very reliable and virtually error free instruments. However if
an error occurs do not attempt to repair the equipment before you understand what caused the error. Note the
problems you have with your instrument and the error messages shown on the display. Then try to solve the
problem according to the error table given below.
Error
Code
Description
Actions to take / Possible cause
FT-15 FT-15D
Under
Weight is too low
X X
Over
Over Load
Check the load;
Load cell or instrument could be defective
X X
Adc Out
Input signal outside the input range
Check the load; Check the calibration;
Load cell or instrument could be defective
X
no Prn
There is no serial interface configured
for printing reports or tickets
Configure the printer interface
X
X
no rCr
No alibi record
Check the numerical entry
X
X
rCr Er
Alibi record data error
Change alibi memory board if this error appears for all CN
X
X
Empty
No data
X
X
ChAnGE Same code already exists in memory
Change vehicle ID code
X
X
Cod FL
Temporary memory is full
Wait for the any 2
nd
weighing cycle or check the memory
X
X
E E E
Load not within zero range
See parameter [202] in chapter 9.4
X
X
Err 1
ADC error respective
Re-energize indicator.If seen again contact FLINTEC.
X
Err 2
Calibration error
Adjust DLC eccentricity; Parameter [320]
X
X
Err 3
Indicator cannot get calibrated
Check load cell and load; Repeat the calibration
X
X
Err 10
EEPROM error
Configure the instrument; EEPROM defective
X
X
Err 20
Calibration error
Calibrate the indicator
X
X
Err 21
Configuration error
Configure the indicator
X
X
Err 22
Tare, CN, Total weight + the SP, PT +
ID in use error
Check SP, PT + ID entries; Check Tare, CN + Total
weight
X X
Err 23
Header Error
Reload the header
X
X
Err 24
Footer Error
Reload the footer
X
X
Err 27
Indicator is not calibrated
Calibrate the indicator
X
X
Err 28
Date & Time error
Adjust Date & Time; replace battery
X
X
Err 29
ID error
Reload ID information
X
X
Err 30
Processor Error
Contact FLINTEC
X
X
Err 34
Indicator cannot get calibrated
Load cell signal is negative, very low or too high
X
X
Err 35
Calibration Error
Calibration test weight is too small;
Check load cell connections
X X
Err 37
Scale unstable
Wait until scale becomes stable;
Check ground connection
X X
Err 40
Alibi memory error
Correct parameter [801]; Alibi memory card broken
X
X
Err 41
New alibi card installed
X X
Err 42
Alibi card serial no. is not identical
Appears when new alibi memory card is installed (*)
X X
Err 43
Alibi card serial no. is not identical, no
print operation will be performed
Identify the alibi memory card to the indicator (*)
X
X
Err 44
Alibi card info error
X X
Err 46
Alibi recording capacity error
X X
Err 47
Info error from weighing board
X
X
Err 50
DLC board not detected
X
Err 61
U10 (eeprom) is not installed
X X
Err 62
U8 (eeprom) is not installed
X X
E XXXX
Hardware error
Re-energize indicator.
If seen again contact FLINTEC.
X X
Ed1 yy ** Communication time out
X
Ed2 yy ** Checksum error
Check DLC (RS-485) hardware; Check DLC cables;
Check DLCs and serial number;
Contact FLINTEC
X
Ed3 yy ** DLC error
Check ambient temperature; Replace DLC
X
Ed4 yy ** DLC is under
X
Ed5 yy ** DLC is over
Check mechanical installation and DLC
X
Ed6 yy ** DLC communication error
Check DLC (RS-485) hardware; Re-energize indicator;
Re-define DLC address
X
Ed7 yy ** DLC address and SN matching error
Check serial number; Re-define DLC address;
Replace DLC
X
Ed0 yy ** DLC status error
Replace DLC
X
* see chapter 9
** yy = number of DLC