FT-15(D) Technical Manual, Rev. 1.02 August 2010

Page 28 of 52

[208 X] Stability Time (for FT-15D only)

If the scale is in no motion during this time period, the scale is regarded as stable. Allowed settings: 0.0….9.9 s

[21-] Scale Build

The capacity and the resolution of the scale will be defined here.

[210 0] Scale Type

The capacity and resolution parameters will vary according to the selection here.

0 = Single Range (default)

1 = Dual Range

You will reach the next parameter by pressing . Please note that the following parameters will vary according
to this selection.

If single range has been selected:

[212 ] Capacity

Press the key to reach this parameter.

[CAP ]

[ XXXXXX ]

The capacity of the scale will be entered here by and keys.
The value will be confirmed by pressing .

[d ]

[ XXXXXX ]

Display resolution of the scale will be selected by key.
The selection will be confirmed by pressing .

If dual range has been selected:

[211 ] 1. Capacity

Press the key to reach this parameter.

[CAP1 ]

[ XXXXXX ]

The capacity of the first weighing range (Select from Appendix 5) will be entered here by
and keys and confirmed with .

[d1 ]

[ XXXXXX ]

Display resolution of the first weighing range will be selected by key and confirmed
with .

[212 ] 2. Capacity

Press the key to reach this parameter.

[CAP2 ]

[ XXXXXX ]

The capacity of the second weighing range (Select from Appendix 5) will be entered here by
and keys and confirmed with .

[d2 ]

[ XXXXXX ]

Display resolution of the second weighing range will be selected by key and
confirmed with .


Attention:

After the scale parameters have been changed the scale has to be calibrated.