13 d, ] diagnostics, ] tests – Flintec FT-15D User Manual
Page 42: ] serial numbers, ] calibration coefficients, ] printing parameter values

FT-15(D) Technical Manual, Rev. 1.02 August 2010
Page 42 of 52
13 D
IAGNOSTICS
The parameter block [9--] contains certain functions for checking and testing.
[9--] Diagnostics
The operations about checking and testing the instrument can be made here.
[90-] Tests
[900 ] Key Pad test
In this step each key’s ASCII code will be shown on the display as you press the related key. By this way you can test
if all the keys are functioning or not. Pressing “
” key will take you to the next parameter.
[901 ] Serial Interface-1 test
The characters in the alphabet will sequentially be transferred from serial interface-1 by pressing the “Zero” key
repeatedly. Received numerical data is shown on display.
[904 ] mV Indication (for FT-15 only)
When pressing the
value is only for test- and service purposes.
[905 ] DLC Internal Count (for FT-15D only)
When pressing the
and down keys and internal counts can be seen by pressing the
Press the
[91-] Serial Numbers
[910]
Instrument Serial No.
[911]
Option 1 Serial No.
[912]
Option 2 Serial No.
[913]
Option 3 Serial No
When pressing the
[97-] Calibration Coefficients
The parameters located in this section are the calibration parameters of the instrument. If these values have
been noted this can help when a calibration problem occurs. You can refresh the calibration by entering the
correct values. Press the
[99-] Printing Parameter Values
[990] Print All Parameters
By pressing the
[991] Loading Default Parameters
When pressing the
the default values or press the