ALTANA Haze-Gard Plus Operating Instructions User Manual
Page 22

delete all
delete last
statistic
on/off
result
on/off
delete all
delete last
C h a n g e C a l V a l u e
I n p u t T r a n s S t d
I n p u t C l a r i t y S t d
S a v e S t d V a l u e s
E s c a p e : S E T U P
T S t d = 1 0 0
1 0 0 0
C o n f i r m : O P E R A T E
E s c a p e : S E T U P
I n i t i a l i z e I n s t r .
O l d N e w
T S t d 1 0 0 1 0 0
C S t d 7 5 . 5 7 5 . 5
S a v e D a t a : O P E R A T E
E s c a p e : S E T U P
Select the
standard you
want to change,
and confirm with
In the four-digit display, the last
number indicates the number of
digits after the decimal point.
Select the
decimal
place to
be
modified.
Change
the value
and
confirm
with
Pressing the
saves the new setting for the stan-
dard value and returns you to the
"Change Calibration Value" subme-
nu. Pressing the
rejects the change and retains the
old standard value.
These new values remain valid only
until the unit is switched off. When it
is switched back on, the old values
are restored.
The "Save Standard Values"
function serves to save the reset
standard values permanently, so
that they are retained even after
the instrument is switched off. If a
change has been confirmed with
sary after exiting from the "Change
Calibration Value" submenu.