ALTANA Haze-Gard Plus Operating Instructions User Manual

Page 22

background image

delete all

delete last

statistic

on/off

result

on/off

delete all

delete last

C h a n g e C a l V a l u e

I n p u t T r a n s S t d
I n p u t C l a r i t y S t d
S a v e S t d V a l u e s

E s c a p e : S E T U P

T S t d = 1 0 0

1 0 0 0

C o n f i r m : O P E R A T E
E s c a p e : S E T U P

I n i t i a l i z e I n s t r .

O l d N e w
T S t d 1 0 0 1 0 0
C S t d 7 5 . 5 7 5 . 5

S a v e D a t a : O P E R A T E
E s c a p e : S E T U P

Select the
standard you
want to change,
and confirm with
.

In the four-digit display, the last
number indicates the number of
digits after the decimal point.

Select the
decimal
place to
be
modified.

Change
the value
and
confirm
with
.

Pressing the button
saves the new setting for the stan-
dard value and returns you to the
"Change Calibration Value" subme-
nu. Pressing the button
rejects the change and retains the
old standard value.

These new values remain valid only
until the unit is switched off. When it
is switched back on, the old values
are restored.

The "Save Standard Values"
function serves to save the reset
standard values permanently, so
that they are retained even after
the instrument is switched off. If a
change has been confirmed with
, a recalibration is neces-
sary after exiting from the "Change
Calibration Value" submenu.