15 error messages – ALTANA Micro-Haze Plus User Manual
Page 45

44
The battery or the
Change or recharge
rechargeable bat-
the battery (chap. 4.1).
tery is discharged.
The capacity of the
If applicable, transfer
memory in STATIS-
the memory contents
TIC MODE is ex-
to a PC and clear the
hausted (999 loca-
memory (chapter 9.3).
tions).
The calibration was
Put unit into the correct
started on the wrong
holder and, if neces-
standard.
sary, restart calibration
(chapter 7).
Inadequate light in-
Repeat calibration. If a
tensity during cali-
fault is detected, please
bration (lamp aging
contact our Customer
or operator error).
Service.
One of the elec-
Please contact our
tronic components is
customer service.
defective.
Lower limit of
Hlog > 10 units min.
measuring range
(chapter 10).
exceeded.
Lower limit of
Gloss > 10 units min.
measuring range
(chapter 10).
exceeded.
15
Error Messages
>>LOW BATTERY!<<
>>LOW BATTERY!<<
>>LOW BATTERY!<<
>>LOW BATTERY!<<
>>LOW BATTERY!<<
MEMORY OVERFLOW
MEMORY OVERFLOW
MEMORY OVERFLOW
MEMORY OVERFLOW
MEMORY OVERFLOW
----->
>>
>
>mode
mode
mode
mode
mode
WRONG HOLDER
WRONG HOLDER
WRONG HOLDER
WRONG HOLDER
WRONG HOLDER
CONTROL LAMP
CONTROL LAMP
CONTROL LAMP
CONTROL LAMP
CONTROL LAMP
UNIT
UNIT
UNIT
UNIT
UNIT
NEEDS SERVICE
NEEDS SERVICE
NEEDS SERVICE
NEEDS SERVICE
NEEDS SERVICE
< 10
< 10
< 10
< 10
< 10
RANGE
RANGE
RANGE
RANGE
RANGE