Specifications – A&D Weighing BM-20 User Manual
Page 82

80
BM
series
17. Specifications
BM-20 BM-22 BM-252
BM-500
BM-300
BM-200
22 g
Weighing capacity
22 g
5.1 g
250 g
520 g
320 g
220 g
22.00008 g
Maximum display
22.000084 g
5.100009 g
250.00084 g 520.0084 g 320.0084 g 220.0084 g
0.01 mg
Minimum display
0.001 mg
0.001 mg
0.01 mg
0.1 mg
0.01 mg
Repeatability
(Standard
deviation) 0.0025 mg
0.004 mg
0.03 mg
0.2 mg
0.1 mg
±0.02 mg
Linearity ±0.010
mg
±0.010 mg
±0.10 mg
±0.5 mg
±0.3 mg
±0.2 mg
Approx. 8 seconds
Stabilization time
(Typical
at
FAST
)
Approx. 10 seconds
Approx. 3.5 seconds
Sensitivity drift
±2 ppm/°C (10 °C to 30 °C, Automatic self calibration: OFF)
Operating environment
5 °C to 40 °C, 85 %RH or less (No condensation)
Display refresh rate
5 times/second or 10 times/second
Environmental sensors
Resolution: Temperature ±1.5 °C, Humidity ±10 %, Air pressure ±10 hPa
Applicable range: 5 °C to 40 °C
Minimum unit
mass
0.1 mg
Counting
mode Number of
samples
10, 25, 50 or 100 pieces
Minimum 100%
reference mass
1 mg
10 mg
Perc
ent
mode Minimum display
0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)
Interface USB
and
RS-232C with Windows Communication Tools Software WinCT
External calibration
weight
20 g
10 g
5 g
2 g
1 g
20 g
10 g
5 g
2 g
1 g
200 g
100 g
50 g
20 g
10 g
500 g
300 g
200 g
100 g
50 g
300 g
200 g
100 g
50 g
200 g
100 g
50 g
Size of loading area for
weighing mm
Weighing pan:
φ25,
Filter weighing:
φ50, φ95
Test tube holder:
φ12
Weighing
pan:
φ90
Test tube holder:
φ12 (Option)
External dimensions
259(W) x 466(D) x 326(H) mm
Power supply &
AC adapter type
Power consumption: Approx. 30VA (supplied to the AC adapter )
Confirm that the adapter type is correct for the local voltage and power
receptacle type.
Mass Approx.
10
kg
The
operating
environment does not include excessive change of ambient temperature, humidity,
vibration, drafts, magnetic fields and static electricity.
The
internal
mass
may change due to corrosion or other damage caused by the operating environment,
or due to aging. Check the internal mass periodically and correct the internal mass value if necessary.
∗1
When placing the tare weight within the standard range, if using the smart range, the weighing sample
can be weighed within the precision range.
∗2
The repeatability for 1 gram within the weighing capacity.
∗3
The repeatability for 100 gram within the weighing capacity.
∗3
∗1
∗2
∗2