Command examples – A&D Weighing BM-20 User Manual
Page 75
BM series
73
15.3.4.
Command Examples
This example uses the "
erCd on
" of "
5if
" so that the
ON command (turning on the balance)
Computer Time
Balance
ON command
Standby mode
Command received
All segments are displayed
Command completed
Zero display
R command (Re-zeroing the display)
Computer Time
Balance
R command
Weighing mode
Command received
In process
Command completed
Zero display
CAL command (Calibration with internal mass)
Computer Time
Balance
CAL command
Zero display
Command received
Command completed
Zero display
All segments