Id number and glp report, Setting the id number – A&D Weighing BM-20 User Manual
Page 54

52
BM
series
11. ID Number And GLP Report
The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The GLP output format is selected at "GLP output (
info
)" of the function table and can be output
to a personal computer or printer using the RS-232C serial interface.
The GLP output format includes the balance manufacturer, model, serial number, ID number,
date, time and space for signature.
The balance can output the following reports for GLP using the RS-232C serial interface.
"Calibration report" of the "Automatic self calibration" and "One-touch calibration" using the
internal mass.
"Calibration
report"
of the calibration using an external weight.
"Calibration test report" of the calibration test using an external weight.
"Title block" and "End block" for the weighing data.
Calibration and calibration test data can be stored in memory and several reports can be output at
the same time. Refer to "
12. Data Memory" for details.
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
Refer
to
"
10.7. Clock and Calendar Function" for checking and storing the time and date.
11.1.
Setting The ID Number
1 Press and hold the
RANGE
key until
ba5fnc
of the function table is displayed, then release
the key.
2 Press
the
RANGE
key several times to display
id
.
3 Press
the
key. Set the ID number using the following keys.
RE-ZERO
key.....The key to set the character of the digit selected.
Refer to the display character set shown below.
RANGE
key........The key to select the digit to change the character.
key..........The key to store the new ID number and display
ba5fnc
.
CAL
key.............The key to cancel the new ID number and display
ba5fnc
.
4 When
displaying
ba5fnc
, press the
CAL
key to return to the weighing mode.