Details of the function table – A&D Weighing BM-20 User Manual
Page 43

BM series
41
10.2.
Details Of The Function Table
Class Item
and
Parameter
Description
0 Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2 Slow response, stable value
SLOW
Can be changed by
response adjustment.
0 Stable range is ±1 digit
1 ±2
digit
#1
5t-b
Stability band width
2 Stable range is ±3 digits
The stabilization indicator
illuminates when the display
fluctuation is within the range.
off OFF
ploc
Display lock function on ON
Display hold function to measure the
volume dispensed from the pipette.
0 OFF
1 Normal
2 Strong
trc
Zero tracking
3 Very strong
Keeps zero display by tracking
zero drift.
0 5 times/second
5pd
Display refresh rate 1 10 times/second
Period to refresh the display
0 Point (.)
pnt
Decimal point
1 Comma (,)
Decimal point format
off OFF
p-on
Auto display-ON on ON
Turns on the weighing mode display
when the AC adapter is connected.
off OFF
ba5fnc
Environment
Display
beep
Beep
on ON
Turns on the weighing mode display
when the AC adapter is connected.
Cl adj
Clock
Refer to "
10.7. Clock and Calendar Function"
The time and date are added to
the output data.
0 Key mode
Accepts the
key only
when the display is stable.
1 Auto print mode A
Reference = zero
2 Auto print mode B
Reference = last stable value
Outputs data when the display is
stable and conditions of ap-p,
ap-b and the reference value are
met.
prt
Data output mode
3 Stream mode /
Interval mode
With data 0, outputs data
continuously; with data 2, uses
interval memory.
0 Plus only
Displayed value > Reference
1 Minus only
Displayed value < Reference
ap-p
Auto print polarity
2 Both
Regardless of displayed value
0 10 digits
1 100 digits
ap-b
Auto print
difference
2 1000 digits
Difference between reference
value and displayed value
#2
0 Not used
1 Stores weighing data
dout
Data output
data
Data memory
2 Stores calibration data
Related items:
prt, int, d-no, 5-td, info
: Factory settings.
#1
The unit of minimum display is digit.
Example: If 1 mg display is selected using the
RANGE
key for the
BM-300, 1 mg is one digit.
#2
Usable minimum display of the balance is one digit.
Example: In gram display, one digit is 0.01 mg for the
BM-252 and 0.1 mg for the BM-300.