Monitoring clock and environmental sensors – A&D Weighing BM-20 User Manual
Page 29

BM series
27
6.6.
Monitoring Clock and Environmental Sensors
In the weighing mode, the built-in clock and environmental sensors (temperature, humidity, air
pressure) can be monitored and temperature data can be output with the function settings.
Specifications Of The Environmental Sensors
Sensor
Resolution Measurement range
Applicable range
Temperature
±1.5 °C 5 °C to 40 °C
Humidity
±10 %
0 % to 100 %
Air pressure
±10 hPa
300 hPa to 1100 hPa
5
°C to 40
°C
Caution
The environmental sensors are built into the case. Therefore, these values are not in accordance
with the environmental condition.
Refer
to
"
17. Specifications" for operating condition of the product.
Operation
Press the
SELECT
key in the weighing
mode. The current time, date and sensor
values (temperature, humidity, air pressure)
are displayed for several seconds in order.
The balance returns to the weighing mode
automatically.
If pressing the
SELECT
key while
displaying current data, the display
proceeds to next data.
If pressing the
CAL
key while
displaying current data, the balance
returns to weighing mode.
Note
If adjusting the built-in clock, refer to "
10.7.Clock And Calendar Function".
If appending time and date to data output, refer to the [Time/Date output (5-td)] of
"
10.2.Details Of The Function Table".
If appending temperature data to the data output, refer to [Temperature output (5-tp)] of
"
10.2.Details Of The Function Table".
Humidity
Air pressure
Temperature
Weighing mode
Time
5 seconds
Date
5 seconds
Environmental sensors
10 seconds
Weighing mode