BM series
21
6.2.
Smart Range For BM-22
The smart range for
BM-22 consists of the standard range and precision range (high resolution).
Smart range function
These ranges can be switched by the weighing value automatically.
When placing the tare weight within the standard range, if pressing the
RE-ZERO
key, the
weighing sample can be weighed within the precision range.
When pressing
RANGE
key, the current range is fixed to the standard range.
Example
1 Press the
RE-ZERO
key to use the precision range.
The balance displays zero.
2 Place the container (tare) on the weighing pan.
When mass of the container (tare) is within the
standard range, the current range is automatically
switched to the standard range.
3 Press the
RE-ZERO
key to use the precision range.
The balance displays zero.
4 Place the weighing sample in the container (tare).
When the sample is within the precision range,
the weighing value can be read in high resolution.
Precision Range And Standard Range For BM-22
Unit Precision
range
Standard
range
Milligram 0.000 to
5100.009
5100.01
to 22000.84
Gram
0.000000
to 5.100009
5.10001
to 22.00084
Refer to page 14 for other units.
Precision range
Standard range
Precision range
Precision range
Container (Tare)
Weighing matter