Calibration report using an external weight – A&D Weighing BM-20 User Manual
Page 57

BM series
55
Calibration report using an external weight
Key operation
1 Press and hold the
CAL
key to display
Cal out
and release the key.
2
Cal 0
is displayed.
3 When updating the calibration mass value, press the
RANGE
key and proceed to step 4.
When using the preset calibration mass value, proceed to step 5.
4 Specify calibration mass value using the following keys.
RANGE
key........The key to select the blinking figure
RE-ZERO
(+)key...The key to increase the value of the blinking figure.
MODE
(-)key .......The key to decrease the value of the blinking figure.
key .........The key to store the new value.
5 When
pressing
the
RE-ZERO
key, the zero point is measured and the weight value is displayed
for a few seconds.
6 Place the displayed mass on the pan. Press the
key to measure the mass and the
weight value is displayed for a few seconds.
7 If GLP output is used,
glp
is displayed and the calibration report is output.
8 The balance returns to weighing mode after this calibration.
Setting
of
"info 1" Setting
of
"info 2"
AD-8121 printer format General
format
Space, ASCII 20h
Terminator, CR , LF or CR
CR
Carriage
return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration weight
Signature
-----------
A
-
&
-
D
MODEL
-----
BM-300
S/N
-----
01234567
ID
------
LAB-0123
DATE
--
2011/02/20
TIME
----
12:34:56
CALIBRATED(EXT.)
CAL.WEIGHT
----
+200.0000
--
g
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-