Function switch and initialization, Permit or inhibit – A&D Weighing BM-20 User Manual
Page 39

BM series
37
9. Function Switch And Initialization
9.1.
Permit Or Inhibit
The balance stores parameters that must not be changed unintentionally.
Example:
Calibration
data for accurate weighing, Data for adapting to the operating environment,
Control data for the RS-232C interface.
The balance is equipped with five switches for the purpose of protecting parameters. Each switch
can select either "permit" or "inhibit". "Inhibit" protects parameters against unintentional operations.
1 Press
the
ON:OFF
key to turn off the display.
2 While pressing and holding the
key and the
RANGE
key,
press the
ON:OFF
key to display
p5
.
3 Press
the
key. Then the balance displays the function switches.
4 Specify the switches using the following keys.
RANGE
key........... The key to select blinking digit.
RE-ZERO
key........ The key to change the parameter for the selected switch.
0 To inhibit changes. (Can not be used.)
1 To
permit
changes. (Can be used.)
key ............ The key to store the new parameter and return to the weighing mode.
CAL
key................ The key to cancel current operation and return to the weighing mode.
Function table
0 To inhibit changes to the function table.
1 To
permit
changes to the function table.
Calibration using the internal mass (One-touch calibration)
0 To inhibit calibration using the internal mass.
1 To permit calibration using the internal mass.
Calibration using the external weight
0 To inhibit calibration using the external weight.
1 To permit calibration using the external weight.
Automatic self calibration (for variation of ambient temperature)
0 To inhibit automatic self calibration.
1 To permit automatic self calibration.
Internal mass correction
0 To inhibit correction.
1 To
permit
correction.