A&D Weighing BM-20 User Manual
Page 61

BM series
59
12.2.
Data Memory For Weighing Data
Features
The balance can store 200 sets of weighing data in memory (if time and date are added, the balance
can store 100 sets). Even if the AC adapter is removed, the data is maintained in the memory.
It is not necessary to connect a printer or personal computer to the balance continually, because
the balance stores the weighing data in memory.
The data memory function does not occupy a printer or personal computer while weighing.
The data in memory can be displayed on the balance for confirmation.
The data in memory can be output at one time to a printer or personal computer.
Comment: Select items for addition to the output data, such as the ID number, data number, time
and date, in the function settings.
Storing the weighing data
Note If
"
Clr
" blinks in the upper left of the display, delete the data in memory.
1 Set the "Data memory (
data
)" parameter to "
1
".
2 Specify the "Time/Date output (
5-td
)" parameter whether time and date is to added or not.
3 The storing mode depends on the "Data output mode (
prt
)" parameter setting.
Four modes are available to store data.
Key
mode...........................When the
key is pressed and the displayed value is stable, the
balance stores the weighing data.
Auto print mode A..............When the displayed value is stable and the conditions of "Auto print
polarity", "Auto print difference" and zero point (reference value) are met,
the balance stores the weighing data.
Auto print mode B..............When the displayed value is stable and the conditions of "Auto print
polarity", "Auto print difference" and last stable data (reference value)
are met, the balance stores the weighing data.
Interval
mode......................Weighing data is stored at an interval specified in "Interval time (
int
)".
Press the
key to start and stop this mode.
Display Symbols For Data Memory
Memory data being displayed
Interval memory
standby indicator
Interval memory
active indicator
The data amount
in memory
Data number of the data
currently displayed
Memory full
Weighing data
being displayed