beautypg.com

Commands, Command list – A&D Weighing BM-20 User Manual

Page 73

background image

BM series

71

15.3.

Commands

15.3.1.

Command List

Note A command has a terminator added, that is specified using "

5if, Crlf

" of the function

table, and is sent to the balance.

Commands to query weighing data
C

Cancels the S or SIR command.

Q

Requests the weighing data immediately.

S

Requests the weighing data when stabilized.

SI

Requests the weighing data immediately.

SIR

Requests the weighing data continuously.

Commands to control the balance

CAL

Same as the

1

CAL

1

key.

Calibration using the internal mass.

OFF

Turns the display off.

ON

Turns the display on.

P

Same as the

ON:OFF

key.

Turing the display on or off.

PRT

Same as the

PRINT

key.

R

Same as the

RE-ZERO

key.

Zero display.

#1

RNG

Same as the

RANGE

key.

Changing the minimum display.

TR

Tares the balance.

Canceling the container’s weight..

#1

TST

Perform calibration test.. Calibration

test

using the internal mass.

U

Same as the

MODE

key.

Changing the unit.

Commands to request data
?ID

Requests the identification number.

?PT

Outputs the tare value.

#1

?SN

Request the serial number of the balance.

?TN

Request the model name of the balance.

Command to set data
PT:∗∗∗.∗∗∗∗ g Sets the tare value.

#1

#1

Command

R assumes the point as zero and sets the display to zero.

Command

TR cancels the tare weight when the displayed value is greater than zero.

The canceled tare value can be read using the command ?PT.

Command

PT:∗∗∗.∗∗∗∗ g sets the tare value digitally.

An example of this command sets a negative target value and fills with the sample until the
display becomes zero. For details, refer to page 75.
Step 1 Place

a

container.

Step 2

Set the display to zero using the R command.

Step 3

Set a target value using the PT: command.