beautypg.com

Checkline WT3-200 User Manual

Page 8

background image

Model WT3-200 Wire Terminal Pull Tester

User’s Guide


7

8

Battery / AC
adapter indicator

Either the AC adapter icon or battery power icon will be shown, depending on
power conditions. Refer to the Power section for details.

10

High / low limit
indicators

Correspond to the programmed set points. Indicator definitions are as follows:

– the displayed value is greater than the upper force limit
– the displayed value is between the limits

– the displayed value is less than the lower force limit

11 Set

points

The programmed force limits. Typically used for pass/fail type testing. 1, 2, or
no indicators may be present, depending on the configuration shown in the
Set Points menu item.


4.2 Controls

Primary

Label Primary

Function

Secondary

Label Secondary

Function

Powers the tester on and off. Press
briefly to power on, press and hold
to power off. Active only when the
home screen is displayed.

ENTER

Various uses, as described in the
following sections.

ZERO

Zeroes the primary reading and
peaks.

(UP)

Navigates up through the menu and
sub-menus.

MENU

Enters the main menu.

ESCAPE

Reverts one step backwards through
the menu hierarchy.

MODE

Toggles between measurement
modes.

(DOWN)

Navigates down through the menu
and sub-menus.

DATA

Stores a value to memory, transmits
the current reading to an external
device, and/or initiates automatic
data output, depending on setup.

DELETE

Enables and disables Delete mode
while viewing stored data.


4.3 Menu navigation basics
Most of the tester’s various functions and parameters are configured through the main menu. To access
the menu press MENU. Use the UP and DOWN keys to scroll through the items. The current selection is
denoted with clear text over a dark background. Press ENTER to select a menu item, then use UP and
DOWN again to scroll through the sub-menus. Press ENTER again to select the sub-menu item.

For parameters that may be either selected or deselected, press ENTER to toggle between selecting and
deselecting. An asterisk (*) to the left of the parameter label is used to indicate when the parameter has
been selected.

For parameters requiring the input of a numerical value, use the UP and DOWN keys to increment or
decrement the value. Press and hold either key to auto-increment at a gradually increasing rate. When
the desired value has been reached, press ENTER to save the change and revert back to the sub-menu
item, or press ESCAPE to revert back to the sub-menu item without saving. Press ESCAPE to revert one
step back in the menu hierarchy until back into normal operating mode.

Refer to the following sections for details about setting up particular functions and parameters.