Checkline WT3-200 User Manual
Page 21

Model WT3-200 Wire Terminal Pull Tester
User’s Guide
20
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Then press ENTER.
7. The display will appear as follows:
Optionally exercise the load cell several times (at full scale, if possible), then press ENTER.
8. After displaying “PLEASE WAIT…” the display will appear as follows:
9. The display will appear as follows:
Apply a weight equal to the full scale of the instrument, then press ENTER.
Remove the load, leave the fixtures in place, then press ZERO.
10. The display will appear as follows:
CALIBRATION
ENSURE NO LOAD
THEN PRESS ZERO
CALIBRATION
GAIN ADJUST
APPLY FULL-SCALE LOAD
200.0 lbF +/-20%
THEN PRESS ENTER
CALIBRATION
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
Optionally exercise
load cell a few times.
THEN PRESS ENTER
CALIBRATION
APPLY LOAD
1 OF 5
ENTER LOAD:
40.0 LBF
THEN PRESS ENTER