background image

Model WT3-200 Wire Terminal Pull Tester

User’s Guide


20

Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Then press ENTER.

7. The display will appear as follows:

Optionally exercise the load cell several times (at full scale, if possible), then press ENTER.

8. After displaying “PLEASE WAIT…” the display will appear as follows:

9. The display will appear as follows:

Apply a weight equal to the full scale of the instrument, then press ENTER.

Remove the load, leave the fixtures in place, then press ZERO.

10. The display will appear as follows:

CALIBRATION


ENSURE NO LOAD

THEN PRESS ZERO

CALIBRATION

GAIN ADJUST

APPLY FULL-SCALE LOAD

200.0 lbF +/-20%

THEN PRESS ENTER

CALIBRATION


Attach necessary
weight fixtures.

THEN PRESS ENTER

CALIBRATION


Optionally exercise
load cell a few times.

THEN PRESS ENTER

CALIBRATION

APPLY LOAD

1 OF 5

ENTER LOAD:

40.0 LBF

THEN PRESS ENTER