Checkline WT3-200 User Manual
Page 23

Model WT3-200 Wire Terminal Pull Tester
User’s Guide
22
The entered calibration point is too close to the previous point.
12 PASSWORDS
Two separate passwords may be set to control access to the Calibration section and to the menu and
other keys. To access the passwords setup screen, select Passwords from the menu. The display will
appear as follows:
12.1 Calibration Password
Select Calibration from the sub-menu. The display will appear as follows:
To set the password, select Enabled, then Set Password. Use the UP and DOWN keys to increment and
decrement the value, from 0 to 9999. When the desired value has been selected, press ENTER, then
ESC to exit the sub-menu.
12.2 Menu Key Password
If enabled, every time the MENU key is selected, a password must be provided. Select Menu Key from
the sub-menu. Follow the same procedure as described in section 10.1.
12.3 Locking Out Other Keys
Other keys may be locked out individually. Select any combination of keys (MODE, ZERO, DATA) by
pressing ENTER in the Passwords sub-menu. Pressing a locked key will prompt the message “KEY
PROTECTED” and then revert to the previous screen.
PASSWORDS
Calibration
Menu Key
Mode Key
Zero Key
Data Key
CALIBRATION PASSWORD
*
Disabled
Enabled
Set Password
(0000 – 9999)
5000
CALIBRATION
LOAD TOO CLOSE
TO PREVIOUS