Model WT3-200 Wire Terminal Pull Tester

User’s Guide


21

Use the UP and DOWN keys to adjust the load value as required. The load values default to
evenly spaced increments, as indicated by the previously entered number of data points. Apply
the calibration load. Then press ENTER.

Repeat the above step for the number of data points selected.

11. After all the calibration points have been completed, the display will appear as follows:

To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select
“EXIT W/O SAVING”.

12. Any errors are reported by the following screens:

Displayed at the start of calibration if a disallowed unit is selected.

Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.

The calibration weight does not match the set value.


LOAD NOT STABLE

PLEASE TRY AGAIN

CALIBRATION

Units must be lbF.

PLEASE TRY AGAIN

PRESS ENTER

CALIBRATION


LOAD TOO LOW

PLEASE TRY AGAIN

CALIBRATION

COMPLETE


SAVE & EXIT

EXIT W/O SAVING