The tester can be calibrated at up to 10 points. Enter the number of calibration points (at least
one point must be selected).

Note: To achieve the accuracy specification of ±0.2%, it is recommended to calibrate the tester at
5 or more evenly spaced increments, such as 40, 80, 120, 160, and 200 lb loads.

3. To escape the Calibration menu at any time, press ESCAPE. The display will appear as follows:

Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will
return to the menu without saving changes.

4. After the number of calibration points has been entered, press ENTER. The display will appear as

follows:

5. Place the tester horizontally on a level surface free from vibration, then press ZERO. The tester

will calculate offsets, and the display will appear as follows:

If failed:

6. The following screen appears after the offsets have been calculated:

CALIBRATION

NOT COMPLETE


CANCEL

EXIT W/O SAVING

CALIBRATION

OFFSET


Sen.Offset Adj.Failed
Ana.Offset Adj.Failed

CALIBRATION

OFFSET

Sen.Offset Adj.Passed
Ana.Offset Adj.Passed

CALIBRATION

OFFSET


Please wait…

CALIBRATION

OFFSET

Place force tester
horizontal.

THEN PRESS ZERO