Checkline WT3-200 User Manual

Page 18

background image

Model WT3-200 Wire Terminal Pull Tester

User’s Guide


17


10.4.2 Auto Output Settings
Select the output type. One, both, or no output types may be enabled. The display will appear as follows:

10.5 Analog Output
This output can be used for chart recorders, oscilloscopes, data acquisition systems, or any other
compatible devices with analog inputs. The output produces ±1 volt at full scale of the instrument. The
polarity of the signal is positive for compression and negative for tension.

10.6 DATA Key Functions
The DATA key can be configured to perform several functions. To configure the DATA key, select DATA
Key
from the menu. The display will appear as follows:


Three options are available:

Selection

Function when pressing DATA

RS232/USB Output

Outputs data via the serial and USB ports

Mitutoyo Output

Outputs data via Mitutoyo (Digimatic) through the serial port

Memory Storage

Stores a reading to memory (refer to the Memory section for details)


Any combination of the above functions may be selected.

Trig. Threshold

Sets the percentage of full scale at which the break detection feature
becomes active. This threshold is provided to ignore peaks that can occur
during sample loading and unloading. Available settings: 5-50%, in 5%
increments.

Auto Zero Delay

Sets the time delay before the primary and peak readings are zeroed.
Available settings: 1-60 sec., in 1 sec. increments between 1-5 sec., in 5
sec. increments between 5-60 sec.

AUTO OUTPUT

SETTINGS


RS232/USB Output

Mitutoyo Output

DATA KEY

*

RS232/USB Output

Mitutoyo Output
Memory Storage