beautypg.com

8 troubleshooting 80 – BUCHI SpeedExtractor E-916 User Manual

Page 80

background image

8 Troubleshooting

80

SpeedExtractor E-916/E-914 Operation Manual, Version D

Blockages may be pushed out under extreme conditions. Therefore an extraction with the following
conditions should free the lines.

• Extraction conditions: temperature 200 °C, pressure 150 bar, 1 cycle, hold 10 min, discharge

5 min, flush with solvent 2 min and flush with gas 5 min. Use the same solvent, cell and vial size as
for the extraction when the blockage appeared.

• If the SpeedExtractor is still blocked after this extraction it will be necessary to identify the

blocked part. The flow test in the service menu allows locating the blockage within the instrument.
See section 8.2.4.

Error 41

Position X clogged: press repeat to

discharge or relieve the pressure

manually.

Repeat

Pressure released?

Control of pressure in

Control of pressure in

Relieve pressure manually of all

clogged positions

One position is deactivated, extraction of

other positions continues

p < 1 bar

Error 42

Error 42

START

START

p > 1 bar

Relieve pressure

manually

End: heater opens

End: heater opens

No error

Error 41

yes

no

Extraction with all posi-

tions

End: heater opens

STATUS

Error 42

A clogging occured during extrantion. Do you want to

open the cell lift? Alternatively, relieve the pressure

manually

Error 29 – 34

Error 41

START

repeat

STATUS

This manual is related to the following products: