8 Troubleshooting

79

SpeedExtractor E-916/E-914 Operation Manual, Version D

More than one position is clogged (Error 41)
Once the SpeedExtractor has shown the error messages 29 – 34, 35 – 40 or 41 and the pressure is
released following the instructions given in the flow charts, it is then necessary to locate the blockage.
To do so, insert empty cells without filter papers, metal frits and plugs in the positions in question.
Start a flush into the collection vials (see section 6.2.6). If solvent flows into the vials, the cell (sample
inside the cell, the filter or frit) was blocked. If solvent does not flow into the vial, the SpeedExtractor is
blocked and further steps are necessary.

Error 35 – 40

Position X clogged: press repeat to

discharge or relieve the pressure

manually.

Repeat

Pressure released?

Control of pressure in

Relieve pressure manually

No wobble position; position is deacti-

vated. Extraction of other positions

p < 1 bar

Error 42

Error 42

START

START

p > 1 bar

Relieve pressure

manually

End: heater opens

End: heater opens

Control of pressure in

Error 35 – 40

yes

no

Extraction with all posi-

tions

End: heater opens

STATUS

STATUS

START

repeat