beautypg.com

4 built-in safety elements and measures – BUCHI SpeedExtractor E-916 User Manual

Page 14

background image

2 Safety

14

SpeedExtractor E-916/E-914 Operation Manual, Version D

2.5.4

Built-in safety elements and measures

• The heating element is equipped with overtemperature protection which is activated at

260 °C ±10 °C.

• The pressure parts are protected by a mechanical pressure control valve which is activated at

200 bar ±20 bar.

• To start a program at least one extraction position must be activated.
• Safety shield sensor: To start an extraction the protection shield must be closed.
• Vial rack sensor: To start an extraction the vial rack must be installed.
• The presence of extraction cells in the heating block is checked in the tightness test at the begin-

ning of each extraction process.

Seismic tie-down
In earthquake-susceptible regions the instrument
should be tied down by the ventilation slot on the
rear of the instrument.

This manual is related to the following products: