4 Description of function

28

SpeedExtractor E-916/E-914 Operation Manual, Version D

NOTE
The START and STOP buttons are used only for the extraction method but have no effect on func-
tions such as preheat, leak test, or flush. All functions apart from extractions are initiated by function
buttons
b.

4.5.2

Main displays of the instrument

A

B

C

D

A The main display shows the most important parameters of the extraction process, such as the

activated positions in the heating block, the maximum pressure at the position valves, the tempera-
ture of the heating block, the number of extraction cycles, as well as the current cycle and the time
remaining to the end of the process. The status and total remaining time are also shown in the main
display. With the function buttons the three main displays STATUS, EXTRACTION and MENU are
accessible.

B STATUS shows a diagram of the lines, valves, solvent reservoirs, extraction cells, and collection vials

of the instrument. The valves open and close according to the stage of the method and the num-
ber of activated positions. As the name already indicates, this menu is very beneficial to get a quick
overview of the current status of operation.

C EXTRACTION is used to edit and save a new method or to open or delete an existing method. It

also includes two functions that are required prior to operation: activation of the extraction positions
(OCCUPIED POSITIONS) and preheating of the instrument to the operating temperature (PRE-
HEAT).

D MENU involves all functions that are not directly involved in an extraction method but are typically

used at start-up, during maintenance, and service, and for product information.