Example of an extraction method

Example: Extraction of 6×5 g sample in 10 mL extraction cells with acetone 50 % and DCM 50 %,
3×5 min at 80 °C and 100 bar.

Procedure:
• Connect the solvent reservoirs to the corre-

sponding ports: a: acetone, b: DCM

• Activate the extraction positions: EXTRAC-

TION  OCCUPIED POSITIONS  YES.

• Flush with solvent (optional): Insert empty

extraction cells in all positions. MENU 
FLUSH  FLUSH INTO WASTE: Flush with
solvent: 120 s; flow rate 50 mL/min; flush
with gas: 180 s; solvent: a: acetone 50 %;
b: dichloromethane 50 %. Remove the
extraction cells.

• Preheat: EXTRACTION  PREHEAT: 80 °C,

OK, ON. The oven starts heating.

• Create method: EXTRACTION  EDIT

METHOD: Temperature 80 °C; pressure
100 bar; vial 60 mL; cell 10 mL; solvent
(see optional FLUSH); number of cycles 3;
cycles: hold 3 min, discharge 2 min, no vial
change (conditions apply for all 3 cycles);
flush with solvent 2 min; flush with gas
3 min.

• When the set temperature is reached, insert

samples and collection vials (60 mL) and
close the protective shield.

• START

6.4.6

Open an existing method

To open an existing method, proceed as follows:

• Select OPEN METHOD in the EXTRACTION menu.

All saved methods (up to 100) are stored in a
numbered table. Select the desired method and
press OK. The name of the method now appears in
the title of the main display.