beautypg.com

4 septum – BUCHI SpeedExtractor E-916 User Manual

Page 67

background image

7 Maintenance

67

SpeedExtractor E-916/E-914 Operation Manual, Version D

a

b

Needles
The needles need to be replaced regularly (i.e. at least
after every 6 months) to reduce the risk of clogging
by particles of the septa or when they are bent. To
do so, remove the covers a and b. Unscrew the
needles with the wrench P/N 053204 and pull them
out from the bottom. Needles are available as sets of
12, P/N 053675. Clogged needles can be cleaned with
thin wire.

NOTE
Make sure that the sloped sides of the two needles of
one position always point away from each other.

7.2.4

Septum

To reduce the risk of clogged needles and EXHAUST
lines, replace the septa of the collection vials regularly,
i.e. at least after every 5 runs.

Septum

Vessel

Corresponding septum

Narrow-necked vials
(60, 240 mL)

qty 100, P/N 049536

Wide-necked vials
(150, 220 mL)

qty 12, P/N 053677

This manual is related to the following products: