2 electrical connections – BUCHI SpeedExtractor E-916 User Manual
Page 34

5 Putting into operation
34
SpeedExtractor E-916/E-914 Operation Manual, Version D
NOTE
The instrument does not have to be operated under
a fume hood but the exhaust must lead into some
kind of ventilation device.
Never hold the instrument on the collection rack or
pump heads to move the instrument. Always use the
handles on the side.
5.2
Electrical connections
NOTICE
Risk of instrument damage by wrong mains supply.
• External mains supply must meet the voltage given on the type plate
• Check for sufficient grounding
NOTE
• External connections and extension lines must be provided with a grounded conductor lead
(3-pole couplings, cord or plug equipment). All used power cords must meet the input power
requirements.