3 Technical data
21
SpeedExtractor E-916/E-914 Operation Manual, Version D
Technical data of the SpeedExtractor
Description
Technical data
Pressure range
50 – 150 bar
Pressure accuracy
±5 bar
Primary pressure nitrogen connection
6 – 10 bar
Flow rate pump
1 – 50 mL/min
Precision flow rate
±2 %
Precision mixer
±2 % (
±5 % for isopropanol)
Extraction cell size
E-916: 10, 20, 40 mL; E-914: 40, 80, 120 mL
Environmental conditions
For indoor use only
Temperature
5 – 40 °C
Altitude
up to 2000 m
Humidity
maximum relative humidity 80 % for temperatures
up to 31 °C, and then linearly decreasing to 50 % at
40 °C
Noise level
<70 dB
This manual is related to the following products: