8 Troubleshooting

89

B-90 Operation Manual, Version B

Table 8-1: Error messages and their remedy (cont.)
043

O

2

sensor short-circuit

Defective sensor, sensor cable or
internal wiring

Contact BUCHI customer service

046

O

2

concentration high

System not leakproof, N

2

-bottle

empty

Test for leakages. Check system
sealings and N

2

-supply.

047

O

2

concentration low

Valve defective

Contact BUCHI customer service

050

B-295 O

2

concentration

too high

System not leakproof, N

2

-bottle

empty

Test for leakages. Check system
sealings and N

2

-supply.

051

Change gas mode

Electrical communication cable
B-290 / B-295 or dongle not
connected

Check cabling or dongle.

058

EEPROM

Contact BUCHI customer service.

Only for operation with the Inert Loop B-295

200

Connection to B-295 interrupted
or disconnected

Defective connection cable or
instrument is not plugged in

Check connection cable for
damage and insert plug.

201

B-295 plugged in while opera-
tion in process

Connection cable plugged in
after switching on the instrument

Restart the system.

203

Pressure alarm not given during
start-up

Pressure switch defective or
contaminated

Contact BUCHI customer service.

204

Failure cooling unit B-295

Contact BUCHI customer service.

8.2

Malfunctions and their remedy

Table 8-2: Malfunctions and their remedy

Malfunction

Possible cause

Remedy

Equipment cannot be switched on

No voltage

Insert mains plug, examine plug for
damage

Fuse has been activated

Reset fuse manually

Peristaltic pump does not deliver

The rollers are not in contact with the
running surface

Elevate the running surface with the
lever. Adjust the lifting height of the
running surface with the hexagonal
hex-wrench from beneath

System does not heat up

Heater cable not plugged in

Plug in heating circuit

Heating is not switched on

Switch on heat

Nominal entry temperature is below
room temperature

Re-select a new inlet temperature

Fuse has been activated

Reset fuse manually

Heater defective

Contact the BUCHI customer service

Faulty gas flow direction or no flow
in the heating system

Check pipe system